CLïR 5400 Ultra High-Purity Water System
Ultra High-Purity Lab Water with UV Oxidation, Ultrafilter, & 0.2 Micron Technology for use with:
– Genetics
– DNA Sequencing
– Polymerase Chain Reaction (PCR)
Designed specifically for genetics applications, the CLïR 5400 delivers 2.0 LPM of Type I, 18.2 Megohm water quality on demand. The water is purified using a staged purification process which includes high-purity ion exchange resins to remove dissolved minerals, ultraviolet light for bacteria sterilization and TOC reduction, and ultrafiltration for pyrogen removal and nuclease free water. As the water exits the system, a final filter removes particulates and bacteria to reach Type I water specifications.
CLïR 5400 meets the challenges of delivering consistent ultra-pure water for critical research and testing in the life sciences. For applications like PCR, DNA Sequencing, Gel preparation, and general molecular biology, the need for RNasefree water becomes a necessity. Conventional methods use Diethyl pyrocarbonate (DEPC) treatment to in activate RNase enzymes. With the use of the Phoenix Genome, those processes are eliminated and contamination is reduced with an on-demand, ultrafiltration process.
Feature
- 2.5 LPM* OF 18 MEGOHM TYPE 1 REAGENT WATER ON DEMAND
- INTUITIVE, USER FRIENDLY TOUCH-SCREEN INTERFACE
- REAL TIME QUALITY MONITORING WITH LEAK DETECTION
- COMPACT DESIGN WITH REMOTE DISPENSE OPTION
- QUICK CHANGE, NO TOOLS CARTRIDGE DESIGN
- LARGE CAPACITY, LOW COST CONSUMABLES CARTRIDGES
- GENUINE RESINTECH MEDIA
*Note: Reported flow rate is typical but can vary depending on supply pressure and system options
Model | 5100 | 5200 | 5300 | 5400 |
Applications | Basic Chemistry, Academic, IC Buffers | HPLC, GC/MS, Trace Organics | Life Science, Cell Culture, Microbiology | DNA Sequencing, PCR, Electrophoresis |
Conditioning Cartridge | • | • | • | • |
Polishing Cartridge | • | • | • | • |
UV Oxidation | • | • | ||
Ultrafiltration | • | • | ||
0.2 micron | • | • | • | • |
Resistivity | 18.2 MΩ-cm | 18.2 MΩ-cm | 18.2 MΩ-cm | 18.2 MΩ-cm |
Bacteria | < 1 cfu/ml | < 1 cfu/ml | < 1 cfu/ml | < 1 cfu/ml |
Endotoxin** | <0.005 EU/ml | <0.005 EU/ml | ||
Particulates | < 0.2 μm filtration | < 0.2 μm filtration | < 0.05 μm filtration | < 0.05 μm filtration |
TOC | < 15 ppb | < 5 ppb | < 15 ppb | < 5 ppb |
RNase | – | – | < 0.01 ng/ml | < 0.01 ng/ml |
DNase | – | – | < 4 pg/μl | < 4 pg/μl |
Dimensions (H x W x D) | 54.8cm x 37cm x 36.5cm | |||
Weight | 22.7 kg dry / 25 kg operating | |||
Inlet Connection | 3/8” O.D. Tubing (9.53 mm O.D. Tubing) | |||
Power Requirements | 120 VAC/ 60 Hz @ 1.0 amp | |||
Pressure | 90 PSIG Max. / 20 PSIG Min | |||
Temperature (Max.) | 100˚F / 30˚C | |||
Flow Rate (Typical) | 0.67 Gallons Per Minute 2.5 Liters Per Minute < 2.0 LPM with Ultrafilter & Endotoxin filter installed |
* Performance based upon RO / DI or Distillation Feed (< 20 uS/cm, < 50 ppb of TOC, 0.2 micron filtration)
** Includes Endotoxin Removal Capsule Filter